For small specific surface area samples, such as battery materials, organic compounds, biological substances, metal powders, and abrasives with low porosity, the static method often encounters challenges in achieving accurate results. This is because the amount of gas adsorption is minimal, making it difficult to measure precisely. In contrast, the dynamic method, especially when equipped with wind-heat assistance, detector thermostats, and a low-temperature cold trap, offers higher accuracy and sensitivity.
The static method relies on measuring the volume of gas adsorbed at different pressure levels. However, for samples with a specific surface area below 10 m²/g, even a small sample weight like 0.5g may result in only 0.1 ml of nitrogen adsorption. The background volume of the sample tube, typically around 3–5 ml, can significantly affect the accuracy of the measurement. To achieve an error margin of less than 3%, the pressure sensor must have an accuracy of at least 0.03%. Unfortunately, most imported pressure sensors have an accuracy of only 0.1%, and those commonly used in surface analyzers are usually around 0.15%. This means that even under ideal conditions, the uncertainty can reach up to 10%.
Increasing the sample size can help reduce errors, but this also increases the background volume, which limits the overall accuracy. Some manufacturers claim that the static method can handle samples with a surface area as low as 0.0001 m²/g, but such claims are not realistic given current technology.
On the other hand, the high-precision dynamic method improves accuracy by enhancing signal strength, reducing noise, and minimizing external interference. Wind-heat assistance and detector thermostats help stabilize the system, while a low-temperature cold trap ensures better signal clarity. For example, with a 0.5g sample of 1 m²/g material, the peak area can be measured with an error margin of less than 2%.
In summary:
1. For small specific surface area samples (below 10 m²/g), it's best to use a dynamic chromatography instrument with wind-heat assistance and a detector thermostat, as these offer higher resolution and sensitivity.
2. For medium-to-large specific surface area samples, both static and dynamic methods can provide accurate results. However, the dynamic method is faster due to its solid standard reference method, while the static method saves time and liquid nitrogen consumption through the multi-point BET method.
3. If you need to measure both surface area and pore size distribution, a static volumetric method is recommended for more reliable and detailed data.
Choosing the right technique depends on the sample type, required precision, and available equipment. Understanding the strengths and limitations of each method ensures more accurate and meaningful results.